Method development A variety of mobile phases were investigated i

Method development A variety of mobile phases were investigated in the development of an HPLC method suitable for the study. These included methanol�Cwater, 75:25 (%, v/v); acetonitrile�Cwater, 75:25; methanol�Cwater, 50:50; methanol�Cwater, 95:5; and acetonitrile�Cphosphate buffer (pH 3.5-6.5), 80:20. The suitability of the mobile phase was kinase inhibitor DZNeP decided on the basis of the sensitivity of the assay, suitability for stability studies, time required for the analysis, ease of preparation, and use of readily available cost-effective solvents. Standard curve The stock solution (500 ��g/ml) was prepared by dissolving an appropriate amount of solid substance of SRT in methanol. The calibration curve Inhibitors,Modulators,Libraries was made using five standard solutions of different concentrations (10, 20, 50, 100, and 200 ��g/ml).

The standard solutions were prepared by diluting an appropriate volume of stock solution with methanol. Each solution was analyzed in triplicate. The peak area values were plotted against the corresponding analyte concentrations to obtain the linear calibration. Validation of method Precision Precision was considered Inhibitors,Modulators,Libraries at two levels, i.e., repeatability and intermediate precision. Inhibitors,Modulators,Libraries Repeatability of sample application was determined as intra-day variation, whereas intermediate precision was determined by carrying out inter-day Inhibitors,Modulators,Libraries variation at three different concentration (20, 60, and 120 ��g/ml) levels in triplicates. Accuracy, as recovery Accuracy was determined by standard addition method. The preanalyzed samples of SRT (20 ��g/ml) were spiked with the extra 0, 50, 100, and 150% of the standard SRT and the mixtures were analyzed by the proposed method.

The experiment was performed in triplicate. The % recovery of samples, relative standard deviation (% RSD), and standard error of mean (SEM) were calculated at each concentration level. Robustness To evaluate HPLC method robustness, a few parameters were deliberately varied. The parameters included variation in percentage of water and methanol Inhibitors,Modulators,Libraries in the mobile phase, flow rate, and column temperature. Each factor selected to examine was charged at three levels. One factor at a time was changed to estimate the effect. Thus, replicate injections (n = 6) of mixed standard solution at three concentration levels were performed under small changes of four chromatographic parameters (factors). Limit of detection (LOD) and quantification (LOQ) The detection limit of an individual analytical procedure is the lowest amount of analyte in a sample that can be detected but not necessarily quantitated as an exact Entinostat value. The quantitation limit of an individual analytical procedure is the lowest amount of analyte in a sample that can be quantitatively determined with suitable precision and accuracy.

Leave a Reply

Your email address will not be published. Required fields are marked *

*

You may use these HTML tags and attributes: <a href="" title=""> <abbr title=""> <acronym title=""> <b> <blockquote cite=""> <cite> <code> <del datetime=""> <em> <i> <q cite=""> <strike> <strong>